1 University of Arizona, Tucson
2 Children's National Medical Center, Washington, DC
Measures of transient evoked otoacoustic emission (TEOAE) reproducibility were obtained for 506 ears of 260 children and young adults. Frequency-specific and whole reproducibility (WR) data were obtained using ILO88 hardware and software and were analyzed with respect to pure-tone threshold and immittance data for the same ears. Reproducibility scores were most robust in the frequency region of 2,000 Hz. In addition, the 2,000-Hz reproducibility score was the most efficient of all measurements (0.85 to 0.91) in separating normal and hearing-impaired ears, regardless of the frequency or frequencies at which the hearing loss occurred. Conductive hearing loss precluded detection of emissions. The present findings suggest that TEOAEs provide useful information in routine clinical practice and that they may be employed to screen for the presence of hearing loss in children and young adults.
Key Words: hearing, loss, otoacoustic, emissions, transient
Submitted on August 18, 1993
Accepted on March 11, 1995
This article has been cited by other articles:
![]() |
Joint Committee on Infant Hearing, American Academy of Audiology, American Academy of Pediatrics, American Speech-Language-Hearing Association, and Directors of Speech and Hearing Programs in State Year 2000 Position Statement: Principles and Guidelines for Early Hearing Detection and Intervention Programs Pediatrics, October 1, 2000; 106(4): 798 - 817. [Full Text] [PDF] |
||||